Features (Patented as a movable measurement device for measuring device airflow and system impedance)

Features (Patented as a movable measurement device for measuring device airflow and system impedance)

Features (Patented as a movable measurement device for measuring device airflow and system impedance) Enables the selection of the optimal fan for a device An optimal fan for a device can be selected by entering accurate measurement results into thermal design simulation software. Compact and lightweight With a compact design and weight of approximately 6 kg, it is portable enough to measure immobile equipment.

Measurement Functions System Impedance:Measurement of the resistance to the flow of air within a device Operating Airflow:Measurement of the actual airflow that passes through a device when a fan is mounted P-Q Performance:Measurement of airflow versus static pressure characteristics* * Performance curve that illustrates the characteristics of a fan for use within a certain system. It shows the relationship between airflow and static pressure.

Specifications

Model no.

9AT2560S-000□(1)

9AT2560A-000□(1)

9AT2560C-000□(1)

Measurement units

Airflow

m3/min

CFM

CFM

Static pressure

Pa

inchH2O

Pa

Measurement range

Airflow

0.20 to 8.00 m3/min (Resolution: 0.01 m3/min)

7 to 282 CFM (Resolution: 1 CFM)

7 to 282 CFM (Resolution: 1 CFM)

Static pressure

0 to 999 Pa (Resolution: 1 Pa)

0 to 4.01 inchH2O (Resolution: 0.01inchH2O)

0 to 999 Pa (Resolution: 1 Pa)

Measurement accuracy

Airflow

±7% of maximum measurable airflow with each nozzle

Static pressure

±10 Pa (0.04 inchH2O) for measurement results < 200 Pa, ±50 Pa (0.20 inchH2O) for measurement results ≥ 200 Pa

Operating environment

Ambient temperature

0 to 40˚C

Humidity

20 to 85% RH (non-condensing)

Display

Data no., Measurement values (airflow, static pressure(2)), Measurement status, Nozzle selection, Measurement mode selection

Interface

Digital output: Included USB serial adapter

Power supply

Input voltage

100 to 240 VAC, 50/60 Hz

Power consumption

260 VA max.

Dimensions

600 (W) × 250 (H) × 250 (D) mm

Mass

Main unit: Approx. 6 kg, Connection duct (including board holder): Approx. 1.5 kg

Included peripherals

1 Set of measurement nozzles, Plastic mounting board (5 pcs / set), Connection duct, AC power cable (2.5 m), USB serial adapter, Instruction manual, Quick start guide, Data viewer software

(1) The AC power plug shape differs with the number in □ of model numbers. AC power plug included in models with 1 in □ is for Japan and North America regions (2 parallel flat pins + a round grounding pin), Input voltage: 100/120 VAC, 50/60 Hz AC power plug included in models with 2 in □ is for Europe region (2 round pins + a female grounding contact), Input voltage: 220 VAC, 50 Hz AC power plug included in models with 3 in □ is for China region (2 angled flat pins + a flat grounding pin), Input voltage: 220 VAC, 50 Hz Product also available without an AC power cable. Model no. 9AT2560S-0000, 9AT2560A-0000, 9AT2560C-0000 (2) Static pressure values are calculated with standard atmosphere as 1013 hPa at 20˚C.

Carrying case Measurement nozzle case included

Model no.

9AT2560-B001 Please add "CS" to the end of the model no. of Airflow Tester in page 1 when ordering Airflow Tester and carrying case as a set. e.g. 9AT2560S-0001CS

Dimensions

705 (W) × 385 (H) × 415 (D) mm

Plastic mounting boards Sized to fit the duct frame. Five boards included with Airflow Tester.

Model no.

9AT2560-P001

Quantity

5 pcs / set

Dimensions

525 (W) × 275 (H) × 4 (D) mm